PPH-1503 is a high-speed and high-precision DC Power Supply with dual range of 15V/3A or 9V/5A. PPH-1503 is exclusively designed to meet low power consumption requirements and users’ great demands of accuracy, speed and resolution of both voltage and current. Circuits are designed with swift response capability to provide a stable voltage output while experiencing load changes. For example, when switching cellular phone from standby to talk mode, the current consumption will be dramatically changed within milliseconds. When load current exceeded current limit, two operational modes: Limit and Trip can be selected. Under Limit mode, PPH-1503 will automatically switch to constant current (CC) mode. Under Trip mode, voltage output will automatically turn off.
PPH-1503 is designed to provide prompt response to the current change for recovering the output voltage at a minimum variation. The high sampling rate allows fast readback in voltage and current measurement to expedite products’ testing speed, which is able to increase the test throughput of production. Sink function can be used to test the battery charger of portable devices. The high precision performance of PPH-1503 provides the measurement capability in low power consumption level. One built-in DVM (Digital Volt-Meter) with input ports is designed to measure any point on DUT while PPH-1503 is outputting voltage and current. Besides, two ports are located on the front and rear panel which users can only select one to perform tests based upon their connection convenience. PPH-1503 is an ideal power source for production lines, R&D laboratories, device inspection, maintenance centers or facilities with the requirements of a swift and precise power supply with DVM.
In the testing of assembling LED backlight module into products such as TV, Monitor, Tablet and NB, the minimum requirement of current resolution should be at least four digits. Different sized LED backlights require different current consumption. Most LED backlights are 150mA, in which, ordinary power supplies could not provide the above-mentioned high current resolution of four digits. However, PPH-1503 provides 0.1mA (equivalent to 4 4/5 digits) current resolution under 5A setting as well as rapid transmission time of both measurement and commands to fully satisfy the testing requirement of current resolution so as to raise the production capacity. To achieve high working voltage requirements, PPH-1503 can be in series with GW Instek GPR-H series power supplies to conduct the testing of high voltage products.
The specifications apply when the PPH-1503 is powered on for at least 30 minutes under +20°C~+30°C.
DC GENERAL |
MEASUREMENT TIME CHOICES |
0.01 ~ 10PLC1,0.01PLC/step |
AVERAGE READINGS |
1~10 |
|
TYPICAL READING TIME2,3 |
31ms |
|
DC VOLTAGE OUTPUT |
OUTPUT VOLTAGE |
0~15V |
OUTPUT ACCURACY |
± (0.05%+10mV) |
|
PROGRAMMING RESOLUTION |
2.5mV |
|
READBACK ACCURACY3 |
± (0.05%+3mV) |
|
READBACK RESOLUTION |
1mV |
|
OUTPUT VOLTAGE RISING TIME |
0.15ms (10% ~ 90%) |
|
OUTPUT VOLTAGE FALLING TIME |
0.65ms (90% ~ 10%) |
|
LOAD REGULATION |
0.01%+2mV |
|
LINE REGULATION |
0.5mV |
|
STABILITY4 |
0.01%+0.5mV |
|
RECOVERY TIME(1000%LOAD CHANGE) |
<40us (<100mV) <80us (<20mV) |
|
RIPPLE AND NOISE5 |
1mV rms(0~1MHz) 8mVpp(20Hz~ 20MHz) |
|
DC CURRENT (23℃±5℃) |
OUTPUT CURRENT |
0 ~ 5A (0 ~ 9V) 0 ~ 3A (9 ~ 15V) |
SOURCE COMPLIANCE ACCURACY |
±(0.16%+5mA) |
|
PROGRAMMED SOURCE RESOLUTION |
1.25mA |
|
READBACK ACCURACY3 |
5A range: ±(0.2%+400uA) 5mA range: ±(0.2%+1uA) |
|
READBACK RESOLUTION |
5A range: 100uA 5mA range: 0.1uA |
|
CURRENT SINK CAPACITY |
0 ~ 5V: 2A 5 ~ 15V:(2A derate 0.1A)/V |
|
LOAD REGULATION |
0.01%+1mA |
|
LINE REGULATION |
0.5mA |
|
STABILITY4 |
0.01%+50uA |
DVM |
INPUT VOLTAGE RANGE |
0 ~ 20VDC |
INPUT IMPEDANCE |
10Ω |
|
MAXIMUM INPUT VOLTAGE |
-3V, +22V |
|
READING ACCURACY3 |
± (0.05%+3mV) |
|
READING RESOLUTION |
1mV |
|
PULSE CURRENT MEASUREMENT |
TRIGGER LEVEL |
5mA ~ 5A, 5mA/step |
HIGH TIME/LOW TIME/AVERAGE TIME |
33.3us to 833ms, 33.3us/step |
|
TRIGGER DELAY |
0 ~ 100ms,10us/steps |
|
AVERAGE READINGS |
1 ~ 100 |
|
LONG INTEGRATION PULSE TIMEOUT |
1S ~ 63S |
|
LONG INTEGRATION MEASUREMENT TIME |
850ms(60Hz)/840ms(50Hz) ~ 60s,or AUTO time 16.7ms/steps(60Hz), 20ms/steps(50Hz) |
|
LONG INTEGRATION TRIGGER MODE |
Rising, Falling, Neither |
|
OVP |
OVP RANGE |
OFF,ON (0 ~ 15.2V) |
RESOLUTION |
10mV |
|
ACCURACY |
50mV |
|
Others |
PROGRAMMING |
IEEE-488.2(SCPI) |
USER_DEFINABLE POWER_UP STATES |
5 sets |
|
REAR PANEL CONNECTOR |
8Pin:output*4, sense*2, DVM*2 |
|
TEMPERATURE COEFFICIENT |
0.1* specification/ ˚C |
|
POWER CONSUMPTION |
150VA |
|
REMOTE/LOCATION CONNECTOR |
USB/GPIB/LAN |
|
RELAY CONTROL CONNECTOR |
150mA/15V 5Voutput, 100mA |
|
Insulation |
Chassis and Terminal |
20MΩ or above (DC 500V) |
Chassis and AC cord |
30MΩ or above (DC 500V) |
|
Operation Environment |
Indoor use, Altitude: ≤ 2000m Ambient temperature: 0 ~ 40°C Relative humidity: ≤ 80% Installation category: II, Pollution degree: 2 |
|
STORAGE Environment |
TEMPERATURE: -20˚C ~ 70˚C HUMIDITY: < 80% |
INPUT POWER |
90-264VAC, 50/60Hz 6 |
Accessories |
CD 8cmUser manual x1, Quick Start manual x1 Test lead GTL-117 x 1, GTL-203A x 1, GTL-204A x 1 |
Dimensions |
222 (W) x 86 (H) x 363 (D) mm |
Weight |
Approx. 4.2kg |
Remarks
|
User Manual (CD)x 1, Quick Start Guide x 1, Power Cord x 1(Region Dependent)
GTL-117 ( Test Lead 10A Maximum) * 1
GTL-204A (Test Lead 10A Maximum) * 1
GTL-203A (Test Lead 3A Maximum) * 1
OPTIONAL ACCESSORIES:
GTL-248 GPIB Cable
GTL-251 GPIB-USB-HS(High Speed)
GTL-246 USB Cable