Q1 | What is the SMA Cable specification in the GKT-008? |
Q2 | Does the GKT-008 near field probe provide antenna factors? |
is the electric field strength of the signal,is the voltage generated at the output port of the antenna.
Q3 | Is it better to use the H and E probes to measure the electric and magnetic fields of the EMI signal separately? |
Q4 | Can GKT-008 be used with other brands of spectrum analyzers? |
Q5 | What is the difference between using a digital oscilloscope's FFT technology and a spectrum analyzer to analyze EMI signals? |
The bandwidth of the oscilloscope and probe |
The bandwidth indicated by the oscilloscope is 3dB bandwidth. For example, an oscilloscope with a bandwidth of 100MHz will have 3dB (about 30%) distortion when measuring a true 100MHz signal. In order to reduce the measurement distortion caused by the bandwidth, the general choice for the oscilloscope bandwidth selection is 5 times the highest frequency of the DUT signal, for example, measuring a 100 MHz signal, it is recommended to use a 500 MHz digital oscilloscope. |
The bandwidth and frequency curves of the oscilloscope and probe |
The EMI regulations must be at least 1 GHz for electromagnetic radiation measurements. Passing or failing is often determined by 1 to 2 dB. In order to avoid false judgement caused by high frequency EMI or harmonic signal errors, it is recommended to use an oscilloscope with sufficient bandwidth, which means that the oscilloscope needs a bandwidth of up to 5 GHz in order to maintain consistent amplitude measurement accuracy. |
In addition to the oscilloscope, the probe used also has the same bandwidth problem. The active probe has a larger bandwidth, but the price may increase by 5 to 10 times. The bandwidth indicated by the passive probe is that the probe needs to be in the 10:1 position, otherwise the bandwidth will decrease. |
Input resistance |
If an EMI near-field probe is used to connect the oscilloscope for replacing the oscilloscope probe, reflection will occur in the high-frequency part because the input impedance of the oscilloscope is generally 1MW. Hence, the signal will be reflected back to the probe end instead of entering the oscilloscope. This situation will vary with frequency. If you really want to use it, you must use an oscilloscope with an input impedance of 50W. |
The number of bits in the ADC is insufficient | ||||||
The general oscilloscope's ADC is 8-bit, and the special-purpose DSO will have ADC with more bits. GW Instek's GSP-9330 is 23-bit by patented technology in small signal positions. The difference in the number of these ADC bits will make a huge difference in the results of the measurements. | ||||||
The lowest range step of the 8-bit ADC, linear scale (left) and logarithmic scale (right) |
||||||
Assume that the oscilloscope's vertical level has a minimum level of 1mv/div and 10 graticules represent 10mV. The 8-bit ADC is used for sampling measurement. The minimum measurable voltage is, After converting to dBuV: This value is already higher than the upper limit value(2) of Class B of the CISPR22 electromagnetic radiation ITE product without considering the antenna factor(1), and since the resolution is 40uV, the following step reading value is 80uV. , after the conversion, the value is 38dBuV, which means that the EMI signal between 32 and 38dBuV cannot be measured. |
||||||
Note:
|
Sampling rate of the oscilloscope ADC |
According to Nyquist–Shannon sampling theorem, the sampling rate of a digital oscilloscope must be greater than or equal to twice the highest frequency of the DUT signal. In the analysis of high-frequency EMI signals, it is more important to select an oscilloscope with a bandwidth of 5 times, and then the sampling rate needs to be more than twice the highest frequency of the DUT signal that requires a high-priced oscilloscope. When GW Instek GSP-9330 is at 10Hz RBW, the base noise is -133dBm when the preamplifier is turned on. That is equivalent to -26dBuV, which measures signals nearly 58dB less than the oscilloscope. Therefore, the difference between the oscilloscope and the spectrum analyzer's ability in measuring small signals is tremendous. |
Q6 | What is the difference between the GSP-9330 EMI Test > Correction > Horizontal/Vertical setting and Sensor Probe > Correction > 3m/10m setting? |
Q7 | Can the GSP-9330 set Trace1 to PK+, Trace2 to QP, and Trace3 to AVG? |
Frequency |
RBW |
<30MHz |
9kHz |
30MHz-1GHz |
120kHz |
>1GHz |
1MHz |
Q8 | Why does the GSP-9330 display the Power Over Range warning message under the EMI Mode? |
Q9 | GSP-9330 can display two Limit Lines of AVG and QP simultaneously in the conduction bandwidth setting of EMC Pretest Mode, but only one can be displayed when self-defining Limit Line. Can two limit lines be displayed? |
Q10 | Can the GSP-9330's built-in EMI test regulations be increased? |
User-defined EMI regulation operating menu | PC software SpectrumShot EMI limit line menu) |
Q11 | How does the GSP-9330 generate an EMI Pretest report? |
Save the measurement result screen on the USB flash drive: Use the USB flash drive to store the measurement results. After the test is completed, open the Peak Table to mark the 10 signals with the highest amplitude, or use the Marker Table to mark up to 6 Markers, and then insert the USB flash drive until the spectrum analyzer identification program is completed, then you can save it. The saved file format is a JPG image file.
The operation procedures are : first insert the USB flash drive on the GSP-9330 front panel USB slot. When the USB icon appears on the screen, press the Quick Save button to complete the screen storage.USB slot and icon of the GSP-9330 (left); The screen can be directly saved on the USB flash drive (right) |
Expressed in MS EXCEL |
Reports generated by SpectrumShot |
This PC software must be collocated with the NI-VISA driver. Please download it from the NI website. |
Q12 | Differences in EMI signal detection using an EMI receiver and GSP-9330 spectrum analyzer. |